2 - 33
(contd)
Alarm
Display Subdivision
5 Test-to-test
compensa-
tion check
For formula No. where
compensated test is
measurable:
(1) The compensation test
must be measurable.
In addition, when the
compensation test is
already registered as a
compensated test in any
preceding formula No.,
the check result for the
formula No. must be
OK.
(2) The compensated test
must be on the right
side of the formula.
(3) When the compensated
test is a photometry
test, no ISE test must
be registered for the
compensation test.
CMP. TEST? 44 to 51
(FORMULA
No.)
This check is not
performed in case
of ORIGINAL ABS.
6 Serum index
analytical
method
check
(1) The assay code for tests
for serum index
measurement must be
RATE-A. This check is
made only when the
tests are measurable.
(2) In the RATE-A test with
serum indexes, no R2
must be specified when
sample blank correction
is carried out.
S. INDEXES? 1 to 37
(CH No.)
7 Cell carry
over
(1) More than 2 types of
carry over evasion (cell)
are specified for 1 test.
(2) Sub-codes (1 to 10)
signify the evasion
types.
CELL. C. O. 1 to 10 Check and correct
the carry over
evasion (cell).
8 ISE carry
over
R1 type is not specified for
ISE test of Reagent Probe
Carry Over Evasion.
ISE C. O. 1 to 40 Check and correct
the Reagent Probe
carry over evasion.
Remarks
Designation
of Check
No.
Details of Check