2 - 19
(cont’d)
Alarm
Control
No.
Alarm Category
Sub-
code
Level Description Remedy
5271 to
5356
SERUM
INDEXES?
104 1 to 36
(ch.)
WARNING (1)Although the sub-code
corresponds to the
serum index
measurement test, the
rate-A assay is not
assigned.
(2)Although the sub-code
corresponds to the
serum index
measurement test and
sample blank is to be
corrected, reagent 2
discharge is specified.
(Analysis does not start.)
Check parameter
for serum indexes.
5431 ON BOARD? 106 1 WARNING There is no measurable
channel.
There is no channel for
which necessary reagent
has been prepared.
(Analysis does not start.)
Check each
parameter.
5441 REAG.POS? 107 1 to 38WARNING (1)The reagent position
specified for a
photometry assay is
also specified for other
photometry assay.
(2)The same reagent
position is specified for
both carry-over cleaning
agent and photometry
assay or ISE test.
(Analysis does not
start.)
Check each
reagent.
5481 CLB(IS)
POS.
108 1 WARNING When ISE is provided,
calibrator or control
positions are set at 55 to 57.
(Analysis does not start.)
Check the position
for control
calibration.
5511 to
5516
ACI ERROR 114 1 to 6WARNING Barcode IC malfunctions.
1:Unassigned
2:Unassigned
3:Sample
4:Unassigned
5:Unassigned
6:Transfer
Replace RSDIST
PC board.