2 - 20
(cont’d)
Alarm
Control
No.
Alarm Category
Sub-
code
Level Description Remedy
5561 to
5568
FD-WRITE? 119 1 WARNING A hardware error has
occurred in writing the
routine sample
measurement data.
119 2 WARNING A hardware error has
occurred in writing the stat
sample measurement data.
119 3 WARNING A hardware error has
occurred in writing the
control sample
measurement data.
119 4 WARNING A hardware error has
occurred in writing the
individual or cumulative
alarm information.
119 5 WARNING A hardware error has
occurred in writing the
parameter data.
119 6 WARNING A hardware error has
occurred in execution of FD
formatting, copying into FD
or FDD cleaning.
119 7 WARNING A hardware error has
occurred in writing the cell
blank data.
119 8 WARNING A hardware error has
occurred in writing the
routine sample test
selecting information.
5571 to
5578
FD READ? 120 1 WARNING A hardware error has
occurred in reading the
routine sample
measurement data.
Refer to alarm
category No. 119.
120 2 WARNING A hardware error has
occurred in reading the stat
sample measurement data.
120 3 WARNING A hardware error has
occurred in reading the
control sample
measurement data.
120 4 WARNING A hardware error has
occurred in reading the
individual or cumulative
alarm information.
(1)Clean the FD.
(2)Replace the FD
with a new one.
(3)Replace the FD
drive.

Table of Contents