2 - 22
(cont’d)
Alarm
Control
No.
Alarm Category
Sub-
code
Level Description Remedy
5631 to
5643
SYSTEM I/F 126 1 WARNING A reception time-out error
has occurred.
126 2 WARNING A transmission time-out
error has occurred.
126 3 WARNING A BCC error or checksum
error has occurred.
126 4 WARNING A parity error has occurred.
126 5 WARNING A framing error has
occurred.
126 6 WARNING An overrun error has
occurred.
126 7 WARNING Frame error
126 8 WARNING Text length error
126 9 WARNING Function character error
126 10 WARNING Sample information error
126 11 WARNING Test selecting information
error
126 12 WARNING Comment information error
126 13 WARNING Reception cannot continue
up to the end code because
an illegal character is
received from the host.
Example)A null code is
received from
the host.
5681 BATH
EXCHANGE
FAILURE
130 1 WARNING The start key has been
pressed despite failure in
incubation bath water
exchange.
Do incubation bath
water exchange
again.
5696 PANEL I/F 140 1 WARNING An error has occurred in
LCD display module
communication.
('Communication error'
appears on the LCD
display.)
(1)Check the
communication
cable of LCD.
(2)Replace the
LCD.
5701 REAGENT? 141 1 to 36WARNING Either reagent positions R1
to R3 or reagent volume
alone is "0" (no
specification).
Check and correct
the contents of
parameters R1 to
R3.
(1)Check the cable
of system I/F.
(2)Check the
contents of
communication
trace.
(3)Check the
contents of
communication
by line analyzer.
(4)Check if a
change has
occurred in
system
parameters.

Table of Contents