2 - 41
(contd)
No. Data Alarm Printer S. I/F Description Remedy
Linearity
abnormal
Number of
photometric
points in reaction
limit level range
9
4 Number of
photometric
points in reaction
limit level range
8
LIN.
LIN. 8
W
F
(Photometry assay only)
In 1-point & rate, rate-A or
rate-B assay, the absorbance
at each photometric point to
be used for calculation of the
absorbance change rate does
not satisfy the following
relational requirement.
E
1
E
b
E
× 100
Linearity limit value
Where, E: Absorbance
change rate determined from
absorbance at each
photometric point in reaction
limit level range by least
squares method
E
1
: Absorbance change
rate in first half
E
b
: Absorbance change
rate in second half
Assuming that the number of
photometric points in the
reaction limit level range is N,
E
1
, E
b
and linearity limit
value can be represented as
shown below.
(1) When N 9
Linearity limit
value LIMIT 8
Make sure the sample
does not contain dust,
etc.
Dilute the sample and
measure it again.
If the stirring motor does
not rotate normally,
report to the
servicemen.
The light source lamp
should not be used
beyond 750 hours.
Replace the light source
lamp and perform cell
blank measurement.
LIMIT8 and LIN.LIMIT
values are both fixed.
(Unit: %,
value: LIN.LIMIT = 10,
LIMIT8 = 30)
E f
E
b