2 - 44
(contd)
No. Data Alarm Printer S. I/F Description Remedy
Sensitivity error SENS Y (Photometry assay)
Sensitivity check is carried out
only for linear (2 to 6-point),
non-linear or isozyme P
calibration.
This error is indicated if the
difference in mean absorbance
between STD (1) and STD (N)
is smaller than SENSITIVITY
LIMIT (input value).
N: Linear (2-point)/
isozyme P calibration ..... 2
Non-linear/linear (3 to 6-
point) calibration ..... 2 to 6
(input value for SPAN
POINT)
Note, however, that N = 2
when CALIB. POINTS = 2 in
non-linear LOGIT-LOG
(3P)/(4P) calibration.
For the mean absorbance of
STD (1) in span calibration, the
following data is used.
Linear ..... Previous S1 ABS
Non-linear ..... Previously
updated mean
absorbance
Sensitivity check is not made
in R.B. calibration.
The parameters on the
screen and FD are not
updated.
Set the standard
solutions and reagent
properly.
Prepare the standard
solutions newly.
Check the sample
syringe for leakage,
clogging, etc.
Unless check is
desired, input 0 for
"sensitivity limit
absorbance" on the
parameter registration
menu.
SD error SD? G (Photometry assay)
This error is indicated upon
completion of non-linear
calibration or if the SD value in
linear (multi-point) calibration is
larger than SD LIMIT (input
value).
Set in correct
concentration series.
Replace the seal piece
of pipettor.
Check fastening of
joints.
Unless check is
desired, input 999.9
for "SD limit
absorbance" on the
parameter registration
menu.