2 - 48
(contd)
No. Data Alarm Printer S. I/F Description Remedy
Sample value
abnormal
R. OVER
&
(ISE)
The sample concentration
(C(S)) is within the following
range.
Na : C(S) < 10.0mEq/L or
C(S) > 250.0mEq/L
K : C(S) < 1.0mEq/L or
C(S) > 100.0mEq/L
Cl : C(S) < 10.0mEq/L or
C(S) < 250.0mEq/L
Dilute the sample and
analyze it again.
Direct measurement is
impossible. Utilize
standard addition
method, etc.
Remove the probe and
correct clogging.
Test-to-test
compensation
error
CMP.T
C
(1) In test-to-test
compensation calculation,
any data alarm other than
shown below is indicated
for the compensation data.
(2) In isozyme Q-channel
concentration calculation,
any data alarm other than
shown below is indicated
for the isozyme P-channel
concentration.
Calculation disabled, test-
to-test compensation
disabled, overflow, random
error, systematic error, QC
error, expected value over
Check the channel data
used for test-to-test
compensation.
Test-to-test
compensation
disabled
CMP.T!
M
(1) In process of calculation for
test-to-test compensation,
the denominator becomes
zero.
(2) The test to be used for test-
to-test compensation is not
measured.
(3) Any test to be used for test-
to-test compensation has
the data alarm of
'calculation disabled' or
'test-to-test compensation
disabled.'
(4) Any compensation test has
the data alarm which
leaves the data blank.
The data is left blank.
Make sure masking is
not specified for the test
to be used for
compensation.
Check the calculation
formula on the
parameter registration
menu 3 screen.
SD absorbance
over
>AMAX
>
At standard concentration of
, absorbance of sample or
absorbance change rate is over
or the same compared with
presumptive absorbance or
absorbance change rate.