Keysight B1505A Configuration and Connection Guide 7-75
Connection and Ordering Examples
Upgrading from existing B1505A
Cables are re-usable in the case listed below
B1513A HVSMU to B1513C HVSMU
B1510A HPSMU to B1511B MPSMU
B1510A HPSMU to B1514A MCSMU
Step 4: Add a N1260A High Voltage Bias-Tee for capacitance measurement up to 3 kV
if necessary
The N1265A Ultra High Current Expander/Fixture does not have a Bias-Tee option
like the N1259A. The N1260A High Voltage Bias-Tee is necessary when making
capacitance measurement up to 3 kV using a N1265A.