Step Front panel Remote interface
1 Press [Home] > User Settings > I/O > LANSettings > Modify
Settings.
Set the first softkey to Manual and press Host Name to enter
the hostname.
SYSTem:COMMunicate:LAN:HOSTname
2 If you change this parameter, you must press [Done] > [Done]
> [Apply Changes] for the change to take effect.
SYSTem:COMMunicate:LAN:UPDate
Each instrument is shipped with a default host name with the format: K-modelnumber-serialnumber,
where modelnumber is the instrument’s 7-character model number (e.g. DAQ970A), and serialnumber is
the last five characters of the 10-character instrument serial number located on the label on the top of the
unit (e.g. 5678 if the serial number is MY12345678).
l The instrument receives a unique hostname at the factory, but you may change it. The host-
name must be unique on the LAN.
l The name must start with letter, other characters can be an upper or lower case letters,
numeric digits, or dashes "-".
l This setting is non-volatile, it will not be changed by cycling power, factory reset (*RST com-
mand), or instrument present (SYSTem:PRESet command).
IP Address
You can enter a static IP address for the instrument as a four-byte integer expressed in dot notation. Each
byte is a decimal value, with no leading zeros (for example, 169.254.2.20). Refer to "More about
IPAddresses and Dot Notation" for further details.
Step Front panel Remote interface
1 Press [Home] > User Settings > I/O > LANSettings > Modify
Settings.
Set the first softkey to Manual and press IP Address to set the
desired IP address.
SYSTem:COMMunicate:LAN:IPADdress
1 Introduction to the Instrument
26 Keysight DAQ970A User's Guide

Table of Contents

Other manuals for Keysight DAQ970A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight DAQ970A and is the answer not in the manual?

Keysight DAQ970A Specifications

General IconGeneral
Scan RateUp to 450 channels/sec
ConnectivityUSB, LAN, GPIB (optional)
Display4.3-inch color LCD
Operating Temperature0 °C to 50 °C
Number of slots3 slots
Measurement TypesDCV, ACV, DCI, ACI, 2-wire resistance, 4-wire resistance, frequency, temperature, capacitance
Voltmeter6.5 digit
Resolution22 bits
MemoryNon-volatile memory