General operation procedure
Lorrca Maxsis User Manual Page 123
Version 5.04 MRN-231-EN
4. Press <F1> or click on NEW...F1 to perform another measurement.
or
Press <
MAIN MENU> to perform other tests (perform in that case also an “END OF DAY WASH (on page
142)”).
or
Exit the software and perform an “E
ND OF DAY WASH (on page 142)”.
7.6. Performing Cell membrane stability test
Start software
7.6.1. Preparing a blood sample
Use the dedicated Lorrca reagents (see Appendix - Parts list (on page 191))
A blood sample can be stored at room temperature during a limited time. After that the sample
cannot be used anymore.
1. Mix 5 ml vial ready to use Elon ISO solution with 25 μl of normal EDTA-anticoagulated blood
(shortly before the measurement).