EasyManua.ls Logo

Lorrca MaxSis - General Options; Elongation; Deformability Curve Parameters

Lorrca MaxSis
229 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Lorrca MaxSis Program
Page 60 Lorrca Maxsis User Manual
Version 5.04 MRN-231-EN
6.4.2. General options
Settings parameter Description Range
Fill empty speed (rps) Cup revolution speed (revolutions per
second) used at the moment of inserting or
removing the test suspension.
[0.1, 10] rps
Shear stress for camera
adjustment (Pa)
Shear stress applied during camera
adjustment (diaphragm).
[0, 100] Pa
Bob temperature (°C) Desired bob (and hence, suspension)
temperature in degrees Centigrade.
[20, 45] °C
Auto start if temp. is stable
within (°C)
If temperature is within ± 0.2 °C, the
measurement starts after the value given
at parameter ... for at least (see next)
[0.1, 10] °C
… for at least (s) Time to automatic start
[5, 300] s
6.4.3. Elongation
6.4.3.1. Deformability curve Parameters
Settings parameter Description
Logarithmic shear stress
axis
Checkbox Turns this option on/off

Table of Contents