Lorrca Maxsis User Manual Page 186
Version 5.04 MRN-231-EN
Work instruction Number 516
Page 1 of 4 Purpose: Preparation for transport
Safety Bio hazard area
Instrument:Lorrca
Version:1, June 2018
RR Mechatronics shall not in any event be liable for any direct, indirect, special, incidental, or
consequential damages, including, without limitation, lost revenues, or lost profits, arising out of or
in any way connected with the use or misuse of the information or lack of information in this
Instruction.
11.1. Preparation for transport.
Perform the preparation for long time inactivity or temporary storage:
In case of long-time inactivity or temporary storage:
1. Perform an End-of-day-wash to clean all tubing.
(User Manual/General operation procedure/End of Day Wash)
2. Disconnect tubing from bottles except the waste line.
(User Manual/General operation procedure/End of Day Wash)
3. Perform a prime all and a cup rinse step to empty all tubing. (to empty and clean all tubes)
4. Clean the outside of the instrument carefully as prescribed.
(User Manual/Maintenance/Cleaning)
5. Remove all liquid bottles, discard or store the contents.
6. Remove all tubes from the pinch valves, the tubes could stick (due to reagents).
(User Manual/Maintenance/Install LORC070902 tubing set 1)
7. Unlock the tube cassettes of the gradient pump to relief the tubing.
(User Manual/Maintenance/Gradient pump tube replacement)