Lorrca MaxSis Program
Lorrca Maxsis User Manual Page 59
Version 5.04 MRN-231-EN
6.4.1. Files
Settings Description
Storage directory Location for storing result files (measurements, diffraction
pattern images). If a "relative" path/name is given, it starts from
the program directory.
Auto export method Method for storing measurement results (as Comma Separated
Values (CSV) files):
<N
O EXPORT>
The results of the last measurement are always available from
file "Auto export file".CSV". This file is overwritten after each
measurement.
<C
ONCATENATE RESULTS TO SINGLE FILE>
With this selection measurements results are always added to a
file named "NameHistory.CSV". Some text editors or
spreadsheet programs cannot read extremely large CSV-files. It
is therefore advisable to backup and delete this file from the
hard disk regularly, before its size exceeds the applicable
maximum.
<E
NUMERATE FILE NUMBER>
With this selection, each measurement is stored in a separate
file with a name being equal to the patient name (as entered in
the "Insert blood sample dialog"). Additional measurements for
the same patient, will be stored with an enumeration number
[0..n]. For example, the first result file of patient Jones is stored
as Jones_0.CSV, the next as Jones_1.CSV and so on.
Auto export file name
Name of the file (with extension ".CSV") that holds the last
measurement results. If the "Auto export method" is set to
"Concatenate results to single file", the name is also used to
create the name of the file containing the measurement history.
Store diffraction
patterns
Checkbox to enable storing diffraction patterns. Only diffraction
patterns are stored that include the actual measurement, i.e.,
not those during camera adjustment.