Lorrca MaxSis Program
Page 66 Lorrca Maxsis User Manual
Version 5.04 MRN-231-EN
6.4.4.4. FSAR options
Settings parameter Description Range
Perform FSAR
procedure
Checkbox, to include the FSAR procedure in the
aggregation measurement. The FSAR
measurement is performed immediately following
the iteration procedure. Note: The acquisition of the
backscatter intensity during the FSAR measurement
is similar to that during the iteration procedure (one
sample per cup revolution). Acquisition parameters
of the iteration procedure (Samples per shear rate,
Maximum deviation, Iteration timeout) also apply to
the FSAR measurement.
-
Low shear rate (1/s) Low shear rate for determining FSARIscglow/I0. The
low shear rate is taken 10/s by default.
[1, 2500] /s
6.4.5. Osmoscan options
This is only available when “Enable Osmoscan” is checked, and Osmoscan option is available on
the instrument
Settings parameter Description Range
Enable Osmoscan Checkbox, Turns this option on/off
-
Osmoscan shear
stress (Pa)
…
[30, 30] (Pa)
Step size (mOsm/kg) …
[5, 5]
(mOsm/kg)
Pump speed (ml/s)
[0.0053,0.0]
(ml/s)
Total scan time (s) …
540 (s)
Total consumption (ml) …
(ml)
Calc. area between: Checkbox, Min or
[75,75]
…and (mOsm/kg)
[450,450]