General operation procedure
Page 142 Lorrca Maxsis User Manual
Version 5.04 MRN-231-EN
7.8.5. Discard of sample
1. Open the hood.
2. Sample is automatically removed by clicking: <remove sample>
3. Remove syringe from the Luer and replace it with a syringe filled with de-ionized water.
4. Press <CLEAN> and slowly flush the connector during rinsing
5. Dry the Bob, Cup and the connector with cotton / paper towel and air (to assure the tube is
free of water)
6. Place Bob in Cup for next measurement
7.9. End of day wash
When the program is shut down a pop-up screen appears with <END OF DAY WASH> message. To
perform an End of day wash:
End-of-day wash procedure:
1. Remove both tubes on the reagent bottles.