General operation procedure
Lorrca Maxsis User Manual Page 143
Version 5.04 MRN-231-EN
2. Mount the reagent tubes to the cleaning tube LORC070905 from the cleaning solution bottle.
3. Place a tube with cleaning solution at sample position.
4. Follow the software instructions.
5. Remove after finishing of the End-of-day wash the cleaning tube from the reagent tubes.
6. Remove the reagent-bottles and store them at cool conditions.
7. Empty waste bottle.
8. Clean and dry bob and cup, drip plate and sample unit plate. Press OK