Maintenance
Lorrca Maxsis User Manual Page 163
Version 5.04 MRN-231-EN
With the pO2 sensor spot removed and the cup sent to the home position, the fiber optic cable is
visible through the sensor spot hole. This alignment is critical to accurately measure the pO2
sensor spot.
10.5. Visual checks
Report irregularities in the logbook.
Check for scratches on surface projection screen or CUP:
No scratches: Continue
Scratches: Contact RR Mechatronics
Check tube attachments.
When loose: re-attach
When damaged: replace
10.6. Lorrca reagents routine
Remove the reagent-bottles after use and store at a cool and dry place (4-6 °C.)
The instrument bottles have to be cleaned after use. Rinse with hot tap water and purge with Ultra
pure water. Drain the bottles as much as possible and let it dry sufficiently.
Do not forget to change the default viscosity value* in case of a new reagent batch at S
ETTINGS
SCREEN
/GENERAL OPTIONS.
Batch information can be found on the Lorrca website.
1. Go to www.lorrca.com .
2. Select "P
RODUCTS".
3. Click on the link for "B
ATCH INFORMATION".
4. On this page all batch information can be found for each Lorrca reagent
5. Click on "C
ERTIFICATE OF ANALYSIS" of the applicable reagent batch number which is mentioned on
the label.
6. Check or change (if needed) the default medium viscosity before the first test sequence in the
'V
ISCOSITY' field in the Sample data screen
* For Osmoscan measurements the mean of the viscosity of the Osmo HIGH and Osmo LOW has
to be set as default medium viscosity.
For the other measurements the viscosity of the Elon ISO and Oxy ISO (used for blood
preparation) has to be set as default medium viscosity.