Lorrca MaxSis Program
Page 76 Lorrca Maxsis User Manual
Version 5.04 MRN-231-EN
6.5.1.2. Rinse gradiënt Osmoscan screen
Screen R
INSE GRADIENT SYSTEM appears:
1. Open the hood.
2. Remove empty sample tube.
3. Place tube with demi-water (+/- 2,5 ml).
4. Press <R
INSE>. to remove sample and rinse the system. Sample is lost and no longer available.
Screen E
MPTY CUP appears.
Press <C
LEAN>
5. Remove empty (rinse) tube.
6. Press OK and the Osmoscan screen appears.
7. Press <F1> or click on N
EW...F1 to perform another Osmoscan.
or
Press <
MAIN MENU> to perform other tests (perform in that case also an “END OF DAY WASH (on page
142)”).
or
Exit the software and perform an “E
ND OF DAY WASH (on page 142)”.