Maintenance
Lorrca Maxsis User Manual Page 175
Version 5.04 MRN-231-EN
10.10.5.1. Gradient pump tube replacement
After 500 measurements/ half year.
Disconnect water and reagent tubing and pull up the sample needle
Lower the BOB and perform two times a P
RIME ALL function
HARDWARE CHECK [ LoRRca IO ] [ Prime all ] [ Start ]
Open the pump drawer and pull the drawer out.