Trouble shooting
Page 152 Lorrca Maxsis User Manual
Version 5.04 MRN-231-EN
9.1. Error messages
Error Description/Check Remedy
Error 1
Error in ambient temperature measuring
circuit. NTC transmits values that are
rejected (<10 °C or >45 °C).
Exit software and switch off the
LoRRca.
Check communication cables to
computer.
Switch on the LoRRca and start the
software.
Error 7
Error in conductivity measuring circuit.
Measured calibration factor too high or
too low. This happens during initiation
and is not depending on attached
measuring cell or conductivity NTC.
Exit software and switch off the
LoRRca.
Check communication cables to
computer.
Switch on the LoRRca and start the
software.
Error 8
Home sensor is not sensed during
initiation of the servomotor.
Exit software and switch off the
LoRRca.
Check communication cables to
computer.
Switch on the LoRRca and start the
software.
Error 9
BOB heating wrong. Heating takes too
much time to reach temperatures above
25 °C.
Exit software and switch off the
LoRRca.
Check communication cables to
computer.
Switch on the LoRRca and start the
software.
Temperature BOB > 44
0
C
Error
25
I2C error. The I2C communication is not
handled well.
Exit software and switch off the
LoRRca.
Check communication cables to
computer.
Switch on the LoRRca and start the
software.
Error
26
Error in braking system. During initiation
the brake did not activate (Servo is
activated and there are encoder pulses)
Exit software and switch off the
LoRRca.
Check communication cables to
computer.
Switch on the LoRRca and start the
software.