Trouble shooting
Lorrca Maxsis User Manual Page 153
Version 5.04 MRN-231-EN
Error
27
Encoder error detected in servo motor
(no pulses during initiation).
Exit software and switch off the
LoRRca.
Check communication cables to
computer.
Switch on the LoRRca and start the
software.
Error
30
Encoder error detected of the gradient
mixer (no pulses during initiation).
Exit software and switch off the
LoRRca.
Check communication cables to
computer.
Switch on the LoRRca and start the
software.
Check in Settings screen (on page
58) that the Osmoscan option
Checkbox is switched OFF.
Error
31
BOB overflow. Overflow sensor did not
give pulses while the gradient mixer or
rinse pump runs.
Exit software and switch off the
LoRRca.
Check communication cables to
computer.
Switch on the LoRRca and start the
software.
Note: For Error codes > 25: Restart the LoRRca MaxSis, if still NOT OK: Call RR Mechatronics see
Service request protocol