Trouble shooting
Lorrca Maxsis User Manual Page 155
Version 5.04 MRN-231-EN
Pressure sensor not available! The pressure sensor was
activated but it is not present on
this device.
Device has to be sent to RRM for
service
Unknown calibration parameter! Unknown calibration parameter in
calibration file.
Wrong calibration file installed or
file is corrupted. Check calibration
file
Unknown calibration mode!
Unknown calibration mode in
calibration file.
Wrong calibration file installed or
file is corrupted. Check calibration
file
Unknown measurement mode! Unknown measurement mode in
calibration file.
Wrong calibration file installed or
file is corrupted. Check calibration
file
Unknown 2nd calibration type! Unknown second calibration type
in calibration file.
Check calibration file