3. Safety Precautions
Please observe the following precautions to ensure patient’s and operator’s safety when
using this system.
DANGER:
Do not use flammable gasses, such as anesthetic gas, oxygen
or hydrogen, or flammable liquids such as ethanol, near this
system and probes, because there is danger of explosion.
WARNING:
1
Do connect the power plug of this system and power
plugs of the peripherals to wall receptacles that meet the
ratings indicated on the rating nameplate. Using an
adapter or multifunctional receptacle may affect the
system grounding performance, and cause the leakage
current to exceed safety requirements.
2
Do connect the video printer to the auxiliary power outlet
of this system. Use the printing cable provided by this
system to connect the printer. Electric shock may result.
3
Connect the grounding cable before turning ON the
system. Disconnect the grounding cable after turning OFF
the system. Otherwise, electric shock may result.
4
For the connection of power and grounding, follow the
appropriate procedures described in this operator’s
manual. Otherwise, there is risk of electric shock. Do not
connect the grounding cable to a gas pipe or water pipe;
otherwise improper grounding may result or a gas
explosion may occur.
5
Before cleaning the system, disconnect the power cord
from the outlet. System failure and electric shock may
result.
6
This system is not water-proof. Do not use this system in
any place where water leakage may occur. If any water is
sprayed on or into the system, electric shock may result.
If water is accidentally sprayed on or into the system,
immediately turn off the power and contact Mindray
Customer Service Department or sales representative.
7
Store and use the transducers carefully. Do not use a
transducer that has a damaged, scratched surface, or
exposed wiring of any kind. Immediately stop using the
transducer and contact Mindray Customer Service
Department or sales representative.