Probes and Biopsy
18-43
Restore the factory default settings
Click [Load Factory], and the position and angle of the guide line will be restored to the factory
default settings.
Exit the Biopsy Verify status
Cli
ck [Exit], and the system will exit the verification status of the guide line.
18.2.7 Removing the Needle-guided Bracket
NGB-004
Hold the probe in the left hand, unscrew the locking nut with the right hand to open
the retaining clamp, and then raise the needle-guided bracket to separate the
locating bulge from the locating grooves.
NGB-005
Hold the probe and the needle-guided bracket; open the grip knob of the
needle-guided bracket.
NGB-006
Metal needle-guided bracket:
(1) Screw out the nut on the guiding block to loose the guiding block and the
needle-guided bracket.
(2) Take the guiding block away in the direction of the needle tail, and then separate
the needle from the probe and the residual parts of the needle-guided bracket.
(3) Screw out the locking nut on the needle-guided bracket to loose the bracket with
the probe.
(4) Hold the probe with one hand, and then separate the needle-guided bracket from
the probe.
Plastic needle-guided bracket:
(1) Remove the guiding block slightly along the direction of the needle’s tail.
(2) Separate the residual part of the needle-guide bracket and the probe from the
needle.