24 Safety Classification
1. According to the type of protection against electric shock:
a) CLASS I EQUIPMENT
2. According to the degree of protection against electric shock:
a) TYPE-BF EQUIPMENT
3. According to the degree of protection against harmful ingress of water:
a) The main unit belongs to IPX0, and the transducers belong to IPX7.
b) Footswitch: 971 SWNOM belongs to IP68.
4. According to the degree of safety of application in the presence of a FLAMMABLE
ANESTHETIC MIXTURE WITH AIR or WITH OXYGEN OR NITROUS OXIDE:
a) EQUIPMENT not suitable for use in the presence of a FLAMMABLE
ANESTHETIC MIXTURE WITH AIR or WITH OXYGEN OR NITROUS
OXIDE
5. According to the mode of operation:
a) CONTINUOUS OPERATION
6. According to the installation and use:
a) MOBILE EQUIPMENT