Probes and Biopsy
18-42
key to cycle through [Large], [Middle], and [Small].
18.2.6 Verifying the Biopsy Guide Line
WARNING:
1 Prior to each biopsy procedure, be sure to verify the guide
line.
2 If the needle is not consistent with the guide line, do not
perform the biopsy procedure.
NOTE: The B-Mode image must be live and active to verify the onscreen biopsy
guideline.
It is recommended to verify the needle path against the biopsy bracket in a sterile container
and sterile water solution.
Click [Verify] in the Biopsy menu to open the [Biopsy Verify] menu.
îš„ Adjust position
Changing the position value can move the biopsy guide line horizontally.
1) Press <Set> to increase the position value.
2) Press [Back] to decrease the position value.
3) Or rotate the [Multifunction] knob to increase or decrease the position value.
îš„ Adjust angle
The method for adjusting guide line angle is the same to that for adjusting guide line position.
îš„ Save verified settings
After the position and angle of the guide line are adjusted, click [Save], and the system will
save the current settings of the guide line.
NOTE:
When the system is restarted, the biopsy guide lines will not be restored to factory
default positions. Altered biopsy guide line positions must be restored back to
factory defaults.