Probes and Biopsy
18-45
NGB-010
(1) Remove the guiding block slightly along the direction of the needle’s tail, and
separate the residual part of the needle-guide bracket and the probe from the
needle.
(2) Remove the support of needle-guided bracket from the probe.
NGB-011
Hold the probe and the needle-guided bracket, and then open the grip knob of the
needle-guided bracket.
NGB-012
Metal/needle detachable needle-guided bracket:
(1) Screw the nut of the guiding block and remove the guiding block slightly along
the direction of the needle’s tail.
(2) Separate the residual part of the needle-guide bracket and the transducer from
the needle.
(3) Screw the pinch nut of the bracket, and remove the needle-guided bracket from
the transducer.
Plastic/needle detachable needle-guided bracket:
(1) Remove the guiding block slightly along the direction of the needle’s tail.
(2) Separate the residual part of the needle-guide bracket and the transducer from
the needle.
(3) Remove the support of needle-guided bracket from the transducer.
18.2.8 Clean and Sterilize the Needle-guided Bracket
Cleaning
1. Wear sterile gloves to prevent infection.
2. Wash the needle-guided bracket with water or soap water to remove all the external
matters. Or, clean the needle-guided bracket with urethane sponge.
3. Wipe off the water on the needle-guided bracket using sterile cloth or gauze after
washing it.
Please follow the instructions in the manual for cleaning.