Guidance and Manufacturer's Declaration
25-2
TABLE 1
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC EMISSIONS
The system is intended for use in the electromagnetic environment specified below. The
customer or the user of system should assure that it is used in such an environment.
EMISSIONS
TEST
COMPLIANCE
ELECTROMAGNETIC ENVIROMENT
GUIDANCE
RF emissions
CISPR 11
Group1
The system
uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The system
is suitable for use in all
establishments including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes
Harmonic
Emissions
IEC61000-3-2
Class A
Voltage
Fluctuations /
Flicker Emissions
IEC61000-3-3
Compliance