WARNING
BEFORE PROCEEDING WITH ANY MAINTENANCE WORK, READ SECTION 1 - HEALTH AND SAFETY.
TKO dÉåÉê~ä=
WARNING
1) BEFORE REMOVING THE EXTERNAL COVERS OF THE SYSTEM, ISOLATE
THE SYSTEM ELECTRICALLY BY SWITCHING OFF THE INCOMING MAINS
SAFETY ISOLATOR (MOUNTED ADJACENT TO THE MACHINE) .
2) READ THE HEALTH AND SAFETY SECTION AT THE FRONT OF THIS
MANUAL.
What may appear to be faults are sometimes simply the failure on the part of the operator to
follow the correct procedures. Before spending time, therefore, searching for a failed
component, ensure that the pre-power-up check has been carried out (see Section 5) and that
all plugs and connectors are secure.
TKP qêçìÄäÉëÜççíáåÖ=ÅÜ~êí=
Note that the selection of faults given for each symptom may not be exhaustive.
Symptom Possible Faults and Checks
No Power to
system
Also, in PC 2000 on
the Pump Control
page, the PSU
Monitor indicator
is coloured red.
1) Incoming mains safety isolator (mounted adjacent to the machine)
switched off.
2) SYSTEM CONTROL ON button not operated. (Located on front panel).
3) EMERGENCY OFF button not released – twist clockwise to release. (Located
on the control panel).
4) External Emergency Off circuit is open circuit. Check that Socket 7 has a
shorting plug fitted. (Located at rear of machine).
WARNING
THE CIRCUIT BREAKERS CB1 AND CB2 ARE LOCATED
WITHIN THE POWER DISTRIBUTION UNIT. DO NOT
REMOVE THE COVER OF THIS UNIT UNLESS YOU ARE
FULLY TRAINED AND AWARE OF THE RISKS
INVOLVED.
5) Circuit breaker CB1 de-energised.
6) Circuit breaker CB2 de-energised. This CB supplies the ±15V dc and 24V dc
power supply unit. Check FUSES 1, 2 and 3 and the ±15V dc and 24V dc
power supply unit’s outputs.
Troubleshooting
Printed: 25-Oct-07, 11:47 Page 7-3 of 12 Issue 1: August 07