EasyManua.ls Logo

Philips Efficia DFM100 - Page 156

Philips Efficia DFM100
220 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4: Repair Internal Assemblies — Main Chassis
146
4 Connect the two spade connectors of the Therapy PCA High Voltage Cable to J8 and J10 on the
Therapy PCA, see Figure 87 and Ta ble 3 0.
5 Install the Main PCA Module. See Assembling the Main Chassis on page 118.
6 Install the Main Chassis. See “Separating Main Chassis and Rear Case on page 113.
7 Close the case. See “Closing the Case on page 149.
To Complete the Replacement
Run Performance Verification and Safety testing as described in the “Performance Verification”
chapter.

Table of Contents

Other manuals for Philips Efficia DFM100

Related product manuals