4: Repair Internal Assemblies — Main Chassis
146
4 Connect the two spade connectors of the Therapy PCA High Voltage Cable to J8 and J10 on the
Therapy PCA, see Figure 87 and Ta ble 3 0.
5 Install the Main PCA Module. See “Assembling the Main Chassis” on page 118.
6 Install the Main Chassis. See “Separating Main Chassis and Rear Case” on page 113.
7 Close the case. See “Closing the Case” on page 149.
To Complete the Replacement
Run Performance Verification and Safety testing as described in the “Performance Verification”
chapter.