Components Troubleshooting 3: Troubleshooting
55
Startup and OpCheck Messages
Tab le 12 lists the messages that can occur at startup or Operational Check.
Table 12 Startup and OpCheck Messages
Message Possible Causes Suggested Solution
Critical failure.
The device cannot deliver
therapy; service immediately.
•Acknowledge
•Run an OpCheck
• Check the HW Error Log and troubleshoot
accordingly
Non-critical failure.
The device can deliver therapy;
service at a nearest opportunity.
Hardware failure
Power Supply or Battery failure Troubleshoot power and battery, see “Power and
Battery” on page 50.
Battery failure
Attempt to import an invalid
Configuration
Reconfigure the device to the user’s settings. See
the Instructions for Use for details.
An error caused the device to
restart
Troubleshoot the device
•
•
Therapy Knob failure during
Op Check
• Make sure the Knob was positioned at 170 J
and 150 J as directed.
• Replace the Therapy Knob
• Replace the Therapy Encoder
error log message
RFU Test is not completed in
time
Run Op Check
error
log message
RFU Test failure • Restart the device
(turn off, wait 10 sec., turn on)
• Replace Processor PCA
error log message
• Restart the device
(turn off, wait 10 sec., turn on)
• Replace SOM PCA
• Restart the device
(turn off, wait 10 sec., turn on
• Replace SOM PCA