Components Troubleshooting 3: Troubleshooting
55
Startup and OpCheck Messages
Tab le 12 lists the messages that can occur at startup or Operational Check.
Table 12 Startup and OpCheck Messages
Message Possible Causes Suggested Solution


Critical failure.
The device cannot deliver
therapy; service immediately.
•Acknowledge
•Run an OpCheck
Check the HW Error Log and troubleshoot
accordingly







Non-critical failure.
The device can deliver therapy;
service at a nearest opportunity.




Hardware failure

Power Supply or Battery failure Troubleshoot power and battery, see “Power and
Battery on page 50.

Battery failure


Attempt to import an invalid
Configuration
Reconfigure the device to the users settings. See
the Instructions for Use for details.



An error caused the device to
restart
Troubleshoot the device


Therapy Knob failure during
Op Check
Make sure the Knob was positioned at 170 J
and 150 J as directed.
Replace the Therapy Knob
Replace the Therapy Encoder






error log message
RFU Test is not completed in
time
Run Op Check


error
log message
RFU Test failure Restart the device
(turn off, wait 10 sec., turn on)
Replace Processor PCA

error log message
Restart the device
(turn off, wait 10 sec., turn on)
Replace SOM PCA


Restart the device
(turn off, wait 10 sec., turn on
Replace SOM PCA