General Service Information
Keep the following points in mind when servicing this product.
Installation
The Efficia DFM100 does not require installation. The Efficia DFM100 Instructions for Use (IFU)
describe the setup required before placing the device into service, as well as configuration options (see
“Passwords” below for the configuration password). All setup activities are designed to be performed by
personnel trained in the proper operation of the product.
To obtain a copy of the Efficia DFM100 Instructions for Use and other Efficia DFM100
documentation in your local language visit:
http://www.philips.com/ProductDocs and follow links to
Browse our document library —> Resuscitation/Defibrillators Products—> Efficia DFM100
(Marketing)—> Instructions for Use.
Display Menus
To display a menu, press the Smart Select knob to bring up the Main Menu. Then turn the Smart
Select knob to scroll up or down through the available choices until the desired selection is highlighted.
To activate the selection, press the Smart Select knob. Select Exit to close the menu without making a
selection.
The menus have a wrap-around scrolling feature - once you reach the bottom of a menu, it
automatically wraps around to the top of the menu and continues scrolling.
Passwords
A password is required in order to access some modes within the defibrillator/monitor. The passwords
are:
• Service Mode: 73275
• Configuration Mode: 752749
Upgrades
Upgrades are available to add specific functionality to the device after purchase. Tab le 1 shows the
available upgrades:
Consult your sales representative, dealer, or distributor for the latest details. See also “Ordering
Supplies and Accessories” on page 177.
Table 1 Efficia DFM100 Upgrades
Upgrade # Upgrade Function
866260 SpO
2
866261 NBP
866262 EtCO
2
866263 AED Mode
866264 External Pacing