5: Performance Verification Performance Verification Procedures
174
Insulation Resistance (Not Mandatory)
Figure 135 to Figure 139 illustrate the Insulation Resistance measurements of Efficia DFM100. For
acceptable test results, see Table 34 “Acceptable Test Results” on page 175
Figure 135 Efficia DFM100 Insulation Resistance Test Diagram - Mains Part and
Protective Earth
Figure 136 Efficia DFM100 Insulation Resistance Test Diagram - Mains Part and
Non-Earthed Accessible Conductive Parts
Figure 137 Efficia DFM100 Insulation Resistance Test Diagram - Mains Part and
Applied Parts