5: Performance Verification Performance Verification Procedures
172
– Single Fault Condition is with AC Mains voltage on Applied Parts (both AC line connections
and earth ground intact)
Should be <
50 A. Record as “kk”.
External Paddles/Pads (IEC type BF)
•Source:
– Normal Condition (both AC line connections and earth ground intact)
Should be < 100 A. Record as “lll”.
– Single Fault Condition (separately open neutral and open earth, each in turn)
Should be < 500 A. Record as “mmm”.
•Sink
– Single Fault Condition (with AC Mains voltage on Applied Parts)
(both AC line connections and earth ground intact)
Should be <
5000 A. Record as “nnn”.