Closing the Case 4: Repair
149
Closing the Case
Preparation
1 Assemble the Front Case. See “Internal Assemblies — Front Case on page 93.
2 Assemble the Rear Case Assembly. See “Internal Assemblies — Rear Case Assembly on page 107.
Assembly
1 Connect the UI PCA connection cable and the LCD Display connection cable. See “Separating the
Front Case Assembly on page 91.
2 Close the Front Case and the Rear Case Assembly. See “Separating the Front Case Assembly on
page 91.
3 Install the six M 4x16 case screws. See “Case Screws on page 91.
4 Install the Handle Assembly. See “Handle Assembly on page 84.
5 Install the Bedrail Mount, if present. See “Bedrail Mount” on page 81.
6 Install accessory pouches, if present. See the Efficia DFM100 Instructions for Use.
To C ompl e t e t h e Re p a i r
Run Performance Verification and Safety testing as described in the “Performance Verification”.