Operational Check 3: Troubleshooting
41
Operational Check Completion
At completion of the Op Check:
The message
Operational Check Passedis displayed if all of the tests pass.
If any test fails, the message Operational Check Failed
is displayed along with one or more of the
following messages, depending upon the nature of the failed functionality:
Service Device
Replace Battery
Replace Pads Cable
Replace Paddles Cable
Replace Therapy Cable
Replace ECG Cable
Fix the problem and successfully run the Operational Check to clear the failure.
Operational Check Report
At the end of an Op Check, depending on the AutoPrint configuration setting, either the Op Check
report is printed, or Print Op Check Report to Check Printer prompt is displayed.
To export to a USB flash drive or print a copy of the report:
Select Exportor Print from the Main Menu.
You may export or print any number of copies while in the Op Check mode. After you exit Op Check,
the Op Check report is no longer available, but you still may find the result of the Op Check in the
Op Check summary (see “Operational Check Summary on page 42).