4: Repair Internal Assemblies — Main Chassis
138
Internal Resistors Module
Preparation
1 Open the Case. See “Opening the Case on page 90.
2 Remove the Main Chassis. See “Separating Main Chassis and Rear Case on page 113.
3 Remove the Main PCA Module from the Main Chassis, See “Disassembling Main Chassis on
page 115.
Removal
1 Remove the four M 3x10 screws with lock washers, see Figure 117.
2 Remove the Internal Resistors Module.
Replacement
1 Install the Internal Resistors Module.
Note the short cable resistor should be placed on left side.
2 Install the four M 3 x 10 screws with lock washers, see Figure 117.
3 Twist the wires and bundle them with a cable tie.
4 Install the Internal Resistors Module connectors to J18 and J19 on the Therapy PCA, see Figure 90
and Tab le 3 0.
5 Install the Main PCA Module. See Assembling the Main Chassis on page 118.
6 Install the Main Chassis. See “Separating Main Chassis and Rear Case on page 113.
7 Close the Case. See “Closing the Case on page 149.
To Complete the Replacement
Run Performance Verification and Safety testing as described in the “Performance Verification”
chapter.