Components Troubleshooting 3: Troubleshooting
65
NBP Monitoring Problems
Test the Efficia DFM100 NBP module if indicated by the troubleshooting, or if there is an uncertainty
about the module performance. Experienced and trained Efficia DFM100 users (e.g. nurses or
biomedical and clinical engineers) may perform the NBP test using the NBP testing service tool. The
training material is included in the kit.
To perform the Efficia DFM100 NBP module tests, you need:
• NBP testing service tool (453564501961)
CAUTION: Only qualified service personnel should perform the NBP testing procedures.
NBP Accuracy Test
To perform the NBP accuracy test:
1 Access the Service Mode Main menu as described in “Accessing Service Mode” on page 15.
2 From the Service Mode Main menu, select NBP
3 The NBP Service screen is displayed (see Figure 42). You may hear a soft, high-pitch tone, this is
normal NBP pump operation.
4 Connect the NBP testing kit to DFM100, see Figure 43.
Figure 42 NBP Service Screen
Figure 43 NBP Accuracy Test Connetion
Container
Loosely fitted
NBP cuff
Manometer
Pump