Performance Verification Procedures 5: Performance Verification
157
Performance Verification Procedures
This section gives instructions for performing inspections and running Performance Verification and
Safety tests on the Efficia DFM100. If desired, you can make copies of the “Test and Inspection Matrix
on page 153 and use it to record results.
The Performance Verification procedures are divided into two levels:
Visual Inspection — examining for damage, wear, contamination
Performance Verification and Safety tests — consist of the following tests and checks:
Service Mode tests (including Operational Check), which consist of running the device in
Service Mode, applying signals, measuring, observing behavior, and recording results.
Functional checks, which consist of running the device in its normal operating mode, applying
signals, measuring, observing behavior, and recording results.
Safety tests, which consist of connecting the Efficia DFM100 to a safety analyzer and measuring
results.
The Performance Tests are sequenced to check more basic functions first, and then build on that to check
more complex functions. We recommend you perform these tests in this sequence.
This section is organized into the following topics:
Visual Inspection . . . . . . . . . . . . . . . . . p. 158
Service Mode Tests . . . . . . . . . . . . . . . . . p. 159
Functional Checks . . . . . . . . . . . . . . . . . p. 163
Safety Tests . . . . . . . . . . . . . . . . . . . p. 169