Introduction 2: Maintenance
19
Maintenance
Introduction
This chapter describes routine maintenance on the Efficia DFM100 defibrillator/monitor.
Most routine maintenance, including periodic operational checks, paper replacement, cleaning, etc. is
performed by the user. Refer to the Instructions for Use for detailed information on these maintenance
procedures.
Service personnel are responsible for the following maintenance:
Software Upgrades . . . . . . . . . . . . . . . . . p. 19
EtCO2 Maintenance . . . . . . . . . . . . . . . . p. 25
Software Upgrades
Perform Software upgrades when:
– Replacing Printer or Printer Assembly
– Replacing Processor, SOM, or Therapy PCA.
– Installing software or changing the device’s Language
– Instructed by Goldway Customer Support or a Service Bulletin.
You may download software from the InCenter. Upload the software on a Philips USB Data Drive
(989803171261) for Efficia DFM100 software upgrade.
NOTE: The RFU indicator may display a “red X” at some point during Software Upgrade. This is normal
behavior, as long as an “hourglass” is displayed after the installation.
To choose proper language and upgrade files:
1 Download the latest revison software from InCenter to local PC.
2 Connect Philips USB Data Drive (989803171261) to local PC.
3 Double click the icon , a pop-up self-extracting window will be displayed. See Figure 20.
Figure 20 Software Self-extracting Screen