Performance Verification Procedures 5: Performance Verification
163
USB Test
Test the USB if indicated by the troubleshooting.
The USB test is safe for the data on the USB drive.
The USB test verifies both the USB flash drive and Efficia DFM100 USB port performance. If in doubt,
test the suspect USB flash drive on different devices or different USB flash drives on the suspect device.
To test the USB:
1 Check the physical integrity of the connection:
a Verify there is no dirt, debris, rust, or liquid spill in the USB port or flash drive.
b Verify there is no physical damage to the USB port or flash drive.
If any of these conditions present, then replace both the I/O PCA and USB flash drive to avoid
damaging other ports and drives.
2 Access the Service Mode Main menu as described in Accessing Service Mode on page 15.
3 From the Service Mode Main menu, select USB Test.
4 Observe the USB Test Passed or USB Test Failed message.
5 Press the Smart Select Knob to acknowledge the Test Done
message.
Functional Checks
The functional checks exercise the basic functions of the defibrillator/monitor. They are intended as a
broad check of the devices performance and are used in conjunction with the Service Mode and Safety
tests to verify the performance of the device.
Perform functional checks with the device in a normal clinical operating mode, not in Service Mode.
TIP: Perform functional checks only for the options installed on your device.
If all elements of a test pass, record that test as a PASS. If there is any failure, begin troubleshooting and
repair as needed.
This section is organized into the following topics:
ECG Check . . . . . . . . . . . . . . . . . . p. 164
SpO
2
Check . . . . . . . . . . . . . . . . . . p. 165
NBP Check . . . . . . . . . . . . . . . . . . p. 165
Defibrillator Measurement Test . . . . . . . . . . . . . . p. 165
Defibrillator Test (AC Power at 200 J) . . . . . . . . . . . . . p. 165
Defibrillator Test (Battery Power at 200 J) . . . . . . . . . . . . p. 166
Defibrillator Charge Cancellation Test . . . . . . . . . . . . . p. 167
Pacer Test . . . . . . . . . . . . . . . . . . . p. 167
Synchronized Cardioversion Test . . . . . . . . . . . . . . p. 168
Paddles Safety Check . . . . . . . . . . . . . . . . p. 168