Performance Verification
Overview
This chapter describes how to verify the performance of the Efficia DFM100 monitor/defibrillator after
repairs are complete. This chapter is organized into the following sections:
Required Testing Levels . . . . . . . . . . . . . . . . p. 151
Verification Test Equipment . . . . . . . . . . . . . . . p. 152
Test and Inspection Matrix . . . . . . . . . . . . . . . p. 153
Performance Verification Procedures . . . . . . . . . . . . . p. 157
Required Testing Levels
The Performance Verification and Safety Tests are intended to verify proper operation of the device
following repair. The level of testing required corresponds to the type of repair performed.
External Repairs/Replacements
External repairs/replacements are those involving the repair or replacement of one or more of the
following items. The key point is that the case has not been opened.
Battery Compartment Cover and Latch
•Bedrail Mount
•Labels
Therapy Knob / Smart Select Knob
Paddle Tray and Plates
•Handle
The following tests are required after an External Repair or Replacement when the case has not been
opened:
Perform the Visual Inspection (see “Visual Inspection on page 158).
Run the Operational Check (see “Operational Check on page 159).
Check the Error Logs for critical errors (see “Error Log Messages on page 44).
After the Therapy Knob replacement only: Run the Controls test (see “Controls Test” on page 160).
After the Paddle Tray and Plates replacement only: Run the Paddle Safety Check (see “Paddles Safety
Check on page 168).