Accessing Service Mode 1: Introduction
15
Accessing Service Mode
WARNING: Be sure that the defibrillator/monitor is not connected to a patient when performing any function in
Service Mode.
NOTE: Make sure that you insert a sufficiently charged battery (at least two LEDs light up) into the device or
connect to AC power when you are performing functions in Service Mode.
To access Service Mode:
1 Tu r n t h e T h e r a p y K n o b t o Monitor.
2 Press the Smart Select knob to bring up the Main Menu.
3 Navigate to Other and press the Smart Select knob.
4 From the Other menu, select Service.
The message appears: Leaving clinical mode. Patient monitoring will be turned off.
5 Navigate to Yes and press the Smart Select knob to confirm the Exit Clinical Mode?
request.
You are prompted to enter a password.
6 Enter the password (73275) by scrolling through the list until the desired digit is highlighted.
7 Press the Smart Select knob to enter each digit.
8 Select Done when you have entered all the digits.
9 Press the Smart Select knob to display the Service Mode Main menu, as shown in Figure 18.
Figure 18 Service Mode Main Menu



Printer Module

NBP Module

Main Processor PCA















Therapy Board


SpO2 Module


EtCO2 Module





