Internal Assemblies — Front Case 4: Repair
105
RFU Indicator
Preparation
1 Open the Case. See “Opening the Case on page 90.
2 Position the Efficia DFM100 display side facing down.
3 Remove the UI PCA. See “UI PCA” on page 102.
Removal
1 Remove the RFU Protective Silicone Cover.
2 Remove the RFU Indicator.
Replacement
1 Install the RFU Indicator into the RFU Protective Silicone Cover.
2 Install the RFU Indicator with the RFU Protective Silicone Cover onto the front case.
3 Install the UI PCA. See “UI PCA” on page 102.
To Complete the Replacement
Run Performance Verification and Safety testing as described in the “Performance Verification”
chapter.
Silicon Buttons
Preparation
1 Open the Case. See “Opening the Case on page 90.
2 Position the Efficia DFM100 display side down.
3 Remove the UI PCA. See “UI PCA” on page 102.
Removal
Remove the three Silicon Button Key Pads:
Lead Select/Patient Category Key Pad (2-Key)
Therapy Key Pad (3-Key)
User interface Key Pad (8-Key).
Replacement
1 Install the Silicon Buttons.
2 Install the UI PCA. See “UI PCA” on page 102.
3 Install the LCD Display Assembly. See “LCD Display on page 95.
4 Close the Case. See “Closing the Case on page 149.
To Complete the Replacement
Run Performance Verification and Safety testing as described in the “Performance Verification”
chapter.