3: Troubleshooting Operational Check
40
Leads ECG Test: Depending on your configuration and attached cables, the test may display:
• Pass/ECG Cable
•Pass/No Cable
• Fail/ECG Cable
• Fail/No Cable
• Fail
If a possible reason of the test failure might be a faulty cable connection, then the following prompt
is displayed at the end of all remaining tests:
Leads ECG Test Failed With Cable.
Disconnect ECG Cable to Rerun Test Without Cable.
Disconnect the ECG cable or acknowledge the Proceed Without Rerunning Test prompt.
Pads/Paddles ECG Test: Depending on your configuration and attached cables, the test may display:
• Pass/Pads
• Pass/External Paddles
•Pass/No Cable
• Fail/Pads
• Fail/External Paddles
• Fail/Internal Paddles
• Fail/No Cable
Battery Test: The test may display:
•Pass
• Fail
• None (if battery is not inserted)
• Low Battery (charge the battery)
SpO
2
Test: Performed only if the SpO
2
option is installed.
The test may display:
•Pass
• Fail
NBP Test: Performed only if the NBP option is installed.
The test may display:
•Pass
• Fail
EtCO
2
: Performed only if the EtCO
2
option is installed.
The test may display:
•Pass
•Fail
• Replace Sensor
• No Sensor Detected
Printer Test: The test may display:
•Pass
•Fail
Among other actions, Printer tests determines if the correct font is installed in the printer. If the
Printer Font is not installed in the printer, but is available in the device memory, then the Printer
Font is installed in the printer, while the message Printer Font Download in Progress is displayed.
Depending on the AutoPrint Opcheck Report configuration setting, a successful Printer Test
(Printer Test passed) displays either the Check Printed Report (set AutoPrint to Yes) or Print
Opcheck Report to Check Printer (set AutoPrint to No) prompt.