Components Troubleshooting 3: Troubleshooting
69
tech. alarm Unexpected module
behavior. Philips asks you to
report this unexpected
software problem
1 Disconnect sensor and reboot.
2 Reconnect sensor.
3 Export error logs and report to the Response
Center.
tech. alarm The calculated value is over
the upper limit (99mmHg)
Check the airway adapter and clean or replace
as necessary
Perform an adapter zero
Replace the sensor if persists
tech. alarm Sensor/Sampling line
unplugged
Re-insert sensor /sampling line.
tech. alarm
tech. alarm Sensor voltage overload Replace the sensor
Replace Processor PCA if persists
 error log message Test is not completed in time Run Op Check
•Replace EtCO
2
Module


error log message
EtCO
2
Op Check
Communication Failure
Check the sensor connection
•Replace EtCO
2
Module
•EtCO
2
Module needs service
Replace Processor PCA

error log message
Service required. Possible
sensors end of useful life
Check the sensor connection
•Replace EtCO
2
Module
•EtCO
2
Module needs service

error log message


error log message
Communications failure.
Flultry sensor or connection
Check the sensor connection
•Replace EtCO
2
Module
•EtCO
2
Module needs service
Replace Processor PCA
Table 23 EtCO
2
Monitoring Problems (Continued)
Symptom Possible Causes Suggested Solution