LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Display Shield Replacement
Section Menu Section Contents Procedures Back Index
203
8
Removing the Display Lens Shield
Refer to Figure 8.17 on p. 202.
To remove the display shield from outside the front case:
1. Remove the four screws (3207367-312) from the display shield.
2. Lift the display shield off of the front case assembly.
Installing the Display Lens Shield
Refer to Figure 8.17 on p. 202.
To install the display shield onto the front case:
NOTE: If new display shield is required, install the Display Shield Repair Kit (MIN 3305431-018) (p. 508).
1. Make sure the display lens surface and shield are completely clean and dust free before installing.
2. Install the shield with four screws (3207367-312); torque to 6.8 in-lb using a T10 bit.
3. If installing a new display lens shield, Label Set 3207318-XXX (see Table 9.14 on page 425) is required. Clean display shield
frame with isopropyl alcohol and install label number 10.