LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
ECG Connector Cable (W07) Replacement
Section Menu Section Contents Procedures Back Index
329
8
Removing the ECG Connector Cable (W07)
Refer to Figure 8.108 on p. 328.
To remove the ECG connector cable (3007991-007) from the rear case (refer to Inside Rear Case Diagrams (p. 228)):
1. Disassemble the case as described in Disassembling the Case (p. 181).
2. Disconnect the ECG cable from the J6 connector on the system PCB (A01).
3. Remove the parameter bezel as described in Removing the Parameter Bezel (p. 311).
4. Remove the four screws (202253-763) securing the support bracket (3205680-001) that holds the ECG connector cable
(3007991-007). Remove the connector, seal and screws. Discard the seal and screws.