3. Installation and setup ABL800 FLEX Operator's Manual
Replacement setup, Continued
Solution
warning
This program allows you to select a solution level value in % below which a
warning will be given.
To select the solution level value for the warning to be activated (default value is
25 %), do the following:
Step Action
1.
Highlight the desired solution bottle: touch the bottle on the screen or
use Next Bottle or Prev. Bottle.
2.
Select the solution warning level (in %), using Increase or Decrease.
3.
Select warning values for solution levels in other bottles in the same
manner.
Press Close to confirm the settings.
4.
Reagents –
Expiration
Warning
This program allows you to set the time for the analyzer to give a warning before
the calibration and cleaning solutions with additives will expire.
To set the time for the warning, do the following:
Step Action
1.
Select time, using the up/down arrow buttons in the "Give warning
before reagents expire" box.
The actual bottles of the analyzer depend on the analyzer type.
Press Close to exit the program and confirm the settings.
2.
Continued on next page