ABL800 FLEX Operator's Manual 3. Installation and setup
Replacement setup, Continued
Solution Level
Forecast
This program allows you to change the solution level forecast so that the analyzer
will predict the solution levels in the bottles more precisely.
It will improve both the graphical presentation on the Replacement Status screen
and the accuracy of the solution low level warning.
To change solution level forecast, do the following:
Step Action
1.
Select the solution bottle, using the Next bottle or Prev. bottle button
(the Rinse solution has been selected on the screen).
2.
Select the true level (in %) with the up and down arrow buttons on the
screen.
The bottom of each solution bottle has markings for solution levels.
Other Solutions
Rinse Solution
You can also remove the solution bottle to check the level, and the
analyzer will not enter Hold mode. However, the Refill will be
performed as soon as the bottle has been mounted back again.
25 %
min.
25 %
min.
The selected setting will be displayed on the Analyzer Status screen.
The screen below is an example.
Continued on next page
3-53